Model:JHH1-CTGD-Ⅱ
Overview
Now, China's coal industry is in-depth study, lasting Publicizing (AQ1029-2007) "coal mine safety monitoring system and instrumentation use management practices", in strict accordance with the provisions and requirements of the standard equipment, high standards of Coal Mine Safety Monitoring System and detection equipment,and do a good job seriously and positively related to management, installation, maintenance, adjustment and other aspects of work.AQ1029-2007 Section 8.3.2 states: "The security monitoring equipment and overhaul before use, you must use the specification requirements by product testing, adjust pass, and run in to the interview can only go down 24 ~ 48h." This provision,coal mine safety monitoring system for ensuring the well run, and thus indirectly protect the normal coal mine production safety is important; order to make this provision to be successfully implemented, the company successfully developed the scientific and technological personnel JHH1-CTGD-Ⅱ type gas detector integrated calibrationTaiwan, the integrated verification platform for various small and medium sized mines in China, can easily and quickly adjusted, test the existing coal mine safety monitoring system in a variety of mine detection sensors and instruments.During the trial in the respective mine got good results, we sincerely welcome the experts, leaders and colleagues on the introduction of continuous improvement, sound advice and recommendations to help us to improve the calibrator more in line with mine siteactual needs and jointly make contributions to mine Safety and Health.
The technical conditions
1 The adjustment of environmental conditions
1.1 Ambient temperature: (15 ~ 35) ℃;
1.2 Relative humidity: <85%;
1.3 Atmospheric pressure: (68 ~ 106) KPa;
1.4 no effect on the surrounding environment should be adjusted to interfering gases;
2 standard gas and the main technical parameters of Glass Rotameter
2.1 calibration gas: within the standard of standard gas or calibration gas;
2.2 Glass Rotameter flow range: 60 ~ 600mL/min;
2.3 Rotameter precision glass level: 2.5;
3 Power and related instruments main technical parameters
3.1 Rated input power: AC220V ± 10%; Frequency 50HZ;
3.2 RXN DC power supply: Output Voltage: DC0 ~ 30V continuously adjustable;
Output current: 0 ~ 2A;
3.3 MB3205 measured three semi-DC power supply DC Digital Ammeter
Full scale ± 1.999A; Resolution: 1mA; input resistance: 0.1Ω;
Accuracy: ± 0.5% +1 character;
3.4 SP10B multifunction counter
Frequency Range: HZ-10MHZ;
Period Range: 100ns-1s;
Counting capacity: 108-1;
Sensitivity: 1HZ-10HZ 40mVrms> 10HZ 20mVrms;
4 Tester of the main technical parameters
4.1 The last adjustment sensor types: ≥ 2 species;
4.2 The last adjustment the number of sensors: ≥ 10 units;
Check table 3.4.3 Size: L × W × H (1400 × 200 × 1800) mm;
Check table 3.4.4 Weight: 125kg;
Note: 10 channels (you can also detect gas sensor 10)
Select a specific gas, according to the standard gas. |
|
|
|
|